GLYPHOSATE HERBICIDE ANALYSIS COLUMNS

Pickering Laboratories cation-exchange Glyphosate column is designed and tested for analysis of Glyphosate and its primary metabolite Aminomethylphosphonic acid (AMPA). This short, isocratic method provides separation of the peaks of interest. Reproducible performance is guaranteed run-to-run and column-to-column. 

 

 

 

 

POST-COLUMN CONDITIONS FOR GLYPHOSATE ANALYSIS:

Reagent 1: Oxidizing reagent—100 μL of 5 % Sodium
                  Hypochlorite in 950 mL of GA116
Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of GA104
Reactor 1: 36 °C, 0.5 mL
Reactor 2: ambient. 0.1 mL
Reagents flow rate: 0.3 mL/min

Detection:

Fluorometer: λex 330 nm, λem 465 nm